   Abstract
  Introduction
  Our Modification
  Discussion/Benefits
   References
   Article Figures

 Article Access Statistics
    Viewed47    
    PDF Downloaded1    

Recommend this journal

 

Previous Article  Table of Contents     
INSTRUMENTS AND EQUIPMENTS
Year :   |  Volume :   |  Issue :   |  Page :
 

Minimising in minimally invasive surgery through the use of a novel and flexible super elastic titanium needle suitable for a 3.5- and 5-mm trocar


 Clinic for Surgery, University Clinical Center Tuzla; Department of Surgery, Faculty of Medicine, University of Tuzla, Tuzla, Bosnia and Herzegovina

Date of Submission10-Mar-2021
Date of Decision14-Apr-2021
Date of Acceptance04-May-2021
Date of Web Publication09-Jun-2021

Correspondence Address:
Samir Delibegovic,
Clinic for Surgery, University Clinical Center Tuzla, Tuzla, Bosnia and Herzegovina. Faculty of Medicine, University of Tuzla, Tuzla
Bosnia and Herzegovina
Login to access the Email id

Source of Support: None, Conflict of Interest: None

DOI: 10.4103/jmas.JMAS_84_21

  Abstract 

The use of smaller ports in surgery is the next step in the evolution of minimally invasive procedures. We present findings, using a novel flexible needle made from a super elastic titanium alloy, which demonstrate that it is possible for a 26- and 30-mm needle to pass through a 3.5- and 5-mm trocar. This new approach results in less trauma and improved cosmetic effect in comparison to the classical 10-mm port. Traditional steps such as handling of the needle holders, loading the needle and placing it at the correct angle and direction, inserting the needle into the tissue and finally safely tying a knot remain the same as with the standard procedure. We propose that this improved type of needle creates a refinement opportunity to replace the classic ones during both laparoscopic and robotic surgeries.


Keywords: Laparoscopic suturing, minimally invasive, needle



How to cite this URL:
Delibegovic S. Minimising in minimally invasive surgery through the use of a novel and flexible super elastic titanium needle suitable for a 3.5- and 5-mm trocar. J Min Access Surg [Epub ahead of print] [cited 2021 Jun 14]. Available from: https://www.journalofmas.com/preprintarticle.asp?id=318099



  Introduction Top


A key advantage of laparoscopic surgery is to improve patients' recovery by minimising abdominal wall trauma. In this regard, the use of smaller ports and instruments is the natural course of evolution of minimally invasive surgery.[1]

Laparoscopic suturing remains indispensable for the execution of surgical procedures in the gastrointestinal tract.[2] Passing a metal needle through a 3.5- and 5-mm trocar instead of the standard 10-mm trocar represents a further evolution of laparoscopic surgery.

We describe here the use of a novel flexible needle, made from a super-elastic titanium alloy, which allows extreme flexibility and elastic deformation so that the 26- and 30-mm needle can pass through a 3.5- and 5-mm trocar (JOST, Belgrade, Serbia). This results in a reduction in trauma to the anterior abdominal wall in comparison with a classic 10-mm trocar, and there is also an improved final cosmetic effect.


  Our Modification Top


The patient in this case gave informed consent and his/her anonymity remained preserved.

We used 2-0 Polydioxanone (JOST PDO, Belgrade, Serbia). The novel needle was made out of titanium alloy Nitinol. It was certified by SZUTEST (the accredited body by European commission for the certification of medical equipment - CE sign for medical devices clas III). Type of needle used was Tapper point, with the curvature of ½ (1/2 circle) and the length of 26 mm.

When inserting the needle into the 3.5- and 5-mm trocar, it was necessary to hold the needle by its metal part and not the thread. Inserting the needle through the 3.5-mm trocar was simple and done in one single action [Figure 1]a and[Figure 1] b.
Figure 1: (a and b) Inserting the needle into a 3.5-mm trocar and removing the needle

Click here to view


Handling the needle holders, loading the needle into its holder, placing it at the correct angle and direction, making an incision into the tissue and finally safely tying the knot were exactly the same as with the standard needle. Here, we present the outcomes using the needle for suturing of a perforated duodenal ulcer [Figure 2]. The final cosmetic effect was also improved in comparison to the standard procedure [Figure 3].
Figure 2: The needle and the needle holder whilst suturing a duodenal ulcer

Click here to view
Figure 3: The final cosmetic effect

Click here to view



  Discussion/Benefits Top


Laparoscopic intracorporeal suturing is one the most difficult procedures in laparoscopic surgery. Suturing and knot tying using the needle with thread continue to be an essential part of all laparoscopic gastrointestinal operations.[3] It requires a 10-mm trocar as a standard.

Whilst it has been described previously that the use of the Clarke ligator (the instrument, knot-pusher) allows for a simpler method in laparoscopic suturing, allowing any size curved needle to be inserted through a 5-mm puncture site,[4] we present here that our novel, super-elastic titanium alloy needle provides flexibility and the ability of the needle to pass through a 3.5- and 5-mm trocar, without any additional instrumentation.

The length of the needle can be variable, for example 26 or 30 mm, and can pass through a 3.5-mm and/or 5-mm trocar. Curvature of the needle is ½ (1/2 circle); however, the degree of the curvature is not associated with passing of the needle through the trocar. This is because the alloy itself is so elastic (bendy), allowing the passage of the needle despite of its curvature. In addition, the operator can arrange the desired form during suturing, and the needle will return to its normal shape, without compromising its sharpness.

The cost of JOST PGLA (Belgrade, Serbia) or JOST PDO (polydioxanone) is 12.95 Euros.[5] The price is 20%–50% higher than other brands currently used. However, if this type of needle was to gain wider acceptance and usage by the surgical teams in different countries, we could expect the price to drop significantly.

Importantly, this method of suturing is also cheaper (more affordable) overall if we compare this method with the Clarke one because it does not require additional instrumentation, thus improving its efficiency.

Minimisation of the trauma to the anterior abdominal wall or the thorax can be achieved by minimising the size of the trocar. The skin incision is smaller in comparison to the use of the standard 10-mm trocar. We also speculate that the post-operative pain may be reduced, because, for example, patients who underwent laparoscopic cholecystectomy with three 5-ram ports had, on average, less incisional pain, spent fewer days on pain medication and took fewer prescription pain relief tablets than patients who underwent the same procedure with standard sized trocars, with two 10-ram ports and one 5-ram port.[6] There is also an association between reduced laparoscopy incisions and reduced pain sensation.[7]

Naturally, any effort to reduce invasiveness and improve cosmesis of laparoscopic surgery must not jeopardise safety.[8] The manufacturer currently produces 2-0 and 3-0 polydioxanone and polyglycolide threads. These can be used for most procedures in the abdomen and thorax.

During robotic surgery, there is a trend towards reducing trocar size from 8 to 5 mm and even 3 mm.[9],[10] The presentation of our novel needle presents an opportunity towards the refinement and minimisation of the diameter of the trocar. We propose that the use of a flexible needle, such as the one described in this report, presents both a refinement and an opportunity for implementation in both laparoscopic surgery and robotic surgery.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
  References Top

1.
Alhashemi M, Almahroos M, Fiore JF Jr, Kaneva P, Gutierrez JM, Neville A, et al. Impact of miniport laparoscopic cholecystectomy versus standard port laparoscopic cholecystectomy on recovery of physical activity: A randomized trial. Surg Endosc 2017;31:2299-309.  Back to cited text no. 1
    
2.
Tang B, Zhang L, Alijani A. Evidence to support the early introduction of laparoscopic suturing skills into the surgical training curriculum. BMC Med Educ 2020;20:70.  Back to cited text no. 2
    
3.
Asao T, Yamaguchi S, Tsutsumi S, Mochiki E, Kuwano H. Experimental results and early clinical experience with an easy method for intracorporeal knot tying using a novel laparoscopic needleholder. Surg Endosc 2007;21:1553-6.  Back to cited text no. 3
    
4.
Reich H, Clarke HC, Sekel L. A simple method for ligating with straight and curved needles in operative laparoscopy. Obstet Gynecol 1992;79:143-7.  Back to cited text no. 4
    
5.
Available from: https://jostsurgical.com/en/catalog/. [Last accessed on 2021 Apr 12].  Back to cited text no. 5
    
6.
Leggett PL, Churchman-Winn R, Miller G. Minimizing ports to improve laparoscopic cholecystectomy. Surg Endosc 2000;14:32-6.  Back to cited text no. 6
    
7.
Barkun JS, Barkun AN, Sampalis JS, Fried G, Taylor B, Wexler MJ, et al. Randomised controlled trial of laparoscopic versus mini cholecystectomy. The McGill Gallstone Treatment Group. Lancet 1992;340:1116-9.  Back to cited text no. 7
    
8.
Talamo G, Sucameli F, Imperatore M, Moggia E, Dova L, Francone E, et al. Is Needlescopic cholecystectomy a safe way to improve mininvasiveness and cosmesis in young female patients? Surg Technol Int 2019;34:129-33.  Back to cited text no. 8
    
9.
Stephan D, Darwich I, Willeke F. First clinical use of 5 mm articulating instruments with the senhance® robotic system. Surg Technol Int 2020;37:63-7.  Back to cited text no. 9
    
10.
Montlouis-Calixte J, Ripamonti B, Barabino G, Corsini T, Chauleur C. Senhance 3-mm robot-assisted surgery: Experience on first 14 patients in France. J Robot Surg 2019;13:643-7.  Back to cited text no. 10
    