Users Online : 1703 About us |  Subscribe |  e-Alerts  | Feedback | Login   
Journal of Minimal Access Surgery Current Issue | Archives | Ahead Of Print Journal of Minimal Access Surgery
           Print this page Email this page   Small font sizeDefault font sizeIncrease font size 
  Search
 
 ¤  Next article
 ¤  Previous article 
 ¤  Table of Contents
  
 ¤   Similar in PUBMED
 ¤  Search Pubmed for
 ¤  Search in Google Scholar for
 ¤Related articles
 ¤   Article in PDF (91 KB)
 ¤   Citation Manager
 ¤   Access Statistics
 ¤   Reader Comments
 ¤   Email Alert *
 ¤   Add to My List *
* Registration required (free)  


 ¤  Abstract
 ¤  Introduction
 ¤  Case report
 ¤  Discussion
 ¤  References
 ¤  Article Figures

 Article Access Statistics
    Viewed5360    
    Printed240    
    Emailed2    
    PDF Downloaded200    
    Comments [Add]    

Recommend this journal

 


 
UNUSUAL CASE
Year : 2005  |  Volume : 1  |  Issue : 2  |  Page : 79-81
 

Breakage of fascial closure device during laparoscopic surgery


Department of Minimal Access Surgery, P. D. Hinduja National Hospital & Medical Research Centre, Veer Savarkar Marg, Mahim, Mumbai 400 016, India

Correspondence Address:
Deepraj S Bhandarkar
India House No 2, Kemps Corner, Mumbai 400 036
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0972-9941.16533

Rights and Permissions

 ¤ Abstract 

Breakage of instruments during laparoscopic surgery is rare. However, when it does occur, locating and retrieving the broken part of the instrument can be cumbersome. Moreover, inability to do so may carry serious medicolegal implications. We report a patient in whom the tip of a fascial closure device broke during laparoscopic surgery. This was located by intraoperative fluoroscopy and retrieved from the extraperitoneal plane via a small incision. The paper discusses the probable factors responsible for breakage of the fascial closure device in our patient and reviews the previously reported cases of the rare complication of breakage of instruments during laparoscopic surgery.


Keywords: Laparoscopic surgery, instrument, breakage


How to cite this article:
Katara A N, Bhandarkar DS, Shah R S, Udwadia T E. Breakage of fascial closure device during laparoscopic surgery. J Min Access Surg 2005;1:79-81

How to cite this URL:
Katara A N, Bhandarkar DS, Shah R S, Udwadia T E. Breakage of fascial closure device during laparoscopic surgery. J Min Access Surg [serial online] 2005 [cited 2019 Jun 19];1:79-81. Available from: http://www.journalofmas.com/text.asp?2005/1/2/79/16533



 ¤ Introduction Top


Breakage of instruments during laparoscopic surgery is rare. However, when it does occur retrieval of the broken part can test the ingenuity of the surgeon. Inability to retrieve a broken part may carry serious medicolegal implications. We report a patient in whom the tip of a fascial closure device (FCD) broke intraoperatively. The likely factors responsible for this adverse event are discussed and the literature reviewed to identify cases of instrument breakage occurring during laparoscopic surgery.


 ¤ Case report Top


A 43-year-old lady was scheduled to undergo laparoscopic biopsy of a lymph nodal mass situated along the superior border of the pancreas suspected to be tuberculous in origin. During laparoscopy, the nodal mass was visualized through the gastrohepatic omentum. A falciform lift[1] was considered necessary to elevate the bulky falciform ligament hanging down onto the operative field. A suture held in an indigenously manufactured FCD was introduced to the left of the falciform through a stab incision. As the FCD carrying a polyglycolic acid suture was pushed through the abdominal wall and a sharp spike was seen entering the abdominal cavity, it became apparent that the jaw mechanism at the tip had broken and had lodged itself along the track. The lymph node biopsy was completed without falciform lift and the specimen was retrieved. At this stage, intraoperative fluoroscopy was utilized to identify the position of the tip, which was found lying just under the entry point of the FCD. A 3-cm skin incision was made, deepened through the subcutaneous tissue and the rectus sheath. The broken tip lodged in the extraperitoneal fat was identified with the help of a haemostat used for probing the wound. This was grasped and retrieved. The fascia was closed with a non-absorbable suture material and all skin incisions were sutured. The patient made an uneventful recovery and was discharged 36 hours later. The event was discussed with the patient and her family postoperatively and documented in the case records. She remains well at follow up six months later.


 ¤ Discussion Top


Intraoperative breakage of instruments is rare during open surgery, probably because of the robust construction of open surgical instruments which lack multiple, movable parts. If an instrument does break during open surgery it is more likely to be noticed promptly and retrieved. Laparoscopic instruments, including the reusable "take-apart" metal variety, have more components as compared to their open counterparts and are more delicate. Be as that may, instrument breakage is uncommon during laparoscopic surgery.

Laparoscopic FCD / suture passer is a thin instrument with a sharp tip. We have regularly used this device for closure of fascia at the site of ports as also for exteriorizing stay sutures without any complication in over 2000 cases. After the adverse incident reported above, we carefully inspected the tips of the Berci FCD (Karl Storz, Tuttlingen, Germany) [Figure - 1] A and also of the indigenous device manufactured locally that was used in our patient [Figure - 1] B. In the Berci device, the lower jaw of the tip is an extension of the body of the instrument, whereas the upper movable jaw functions on a hinge. In the device locally manufactured device we had used, the tip consisting of two jaws was detachable and appeared to have been soldered on to the shaft. The FCD, both the Berci and indigenous variety, have a groove in which the suture must be placed. If the suture is not placed in the groove, the profile of the tip is not a smooth tapering point. Pushing such a tip through tissues may require use of excessive force to be applied leading to breakage of the tip. It is unclear whether in our case improper positioning of suture in the groove of the FCD was a factor responsible for breakage of its tip.

Almost all laparoscopic instruments are introduced into the abdominal cavity through ports and thus their tips do not meet with any degree of resistance. In our patient, introduction of the FCD percutaneously through a stab incision may have resulted in generation of excessive force that caused it to break at its weakest point - where the tip was soldered on to the shaft. During arthroscopic procedures, thin instruments with delicate, sharp tips are introduced percutaneously to manipulate intra-articular structures, and breakage of the tip of these instruments is a well-recognised complication.[2] To guard against this eventuality in laparoscopic surgery, it is imperative to check that delicate instruments such as a FCD do not have soldered parts at the tip which make them prone to breakage. Furthermore, if during manufacture or repair of a laparoscopic instrument the inexpensive lead-based or silver solder is used, it weakens the instrument because it is applied at high temperatures. After the use of such solder, the instrument is more likely to break.[3] Repeated autoclaving of reusable instruments may weaken them and make them prone to breakage. Routine and thorough checking of the instruments by those responsible for their maintenance can help avoid intraoperative breakage.

Recently, Salameh reported two cases of breakage of the tip of Gore suture passer (W. L. Gore, & Associates, Newark, DE) during laparoscopic repair of ventral hernia.[4] This author points out that a common mechanism of breakage of suture passer is its withdrawal in a partially open position such that the tip catches on the fascia and breaks. Also, changing the direction of the suture passer half way through its introduction is liable to generate opposing shear forces on the tip resulting in its breakage. Thus, both these points need to be guarded against when using a suture passer to avoid breakage of its tip.

If an instrument breaks during laparoscopic surgery and the broken part lies in the operative field, it can be retrieved immediately. For example, Lynch et al reported the breakage and recovery of a 2-mm segment of needle from an Autosuture Endostitch device (U.S. Surgical) during a laparoscopic Burch procedure which was retrieved uneventfully.[5] Often, however, the breakage becomes apparent after some time or the broken part migrates in the abdominal cavity, making its retrieval difficult. Use of intraoperative fluoroscopy, as was done in our case, is an option to locate and remove the broken instrument part. Kandioler-Eckersberger et al describe the use of a novel method to identify the tip of a reusable laparoscopic grasper which was lost between loops of bowel in two cases.[6] After initial unsuccessful attempts at retrieval by patient positioning, fluoroscopic localization, endoscopic visualization etc, they used a magnetic probe to retrieve the broken parts. The probe consisted of a 6-cm magnetic tip attached to a 40-cm semi flexible Teflon rod. The probe was passed through one of the ports and was placed in the vicinity of the lost part under fluoroscopic guidance; the lost metallic piece was attracted to the magnet. It was retrieved and a laparotomy was avoided.

The action that needs to be taken after an adverse event involving breakage of a medical device varies from hospital to hospital. It is essential to document the event in the operation records and report it to the hospital authority. In the Western countries it may also have to be reported to the relevant state authority in prescribed forms.[7] Discussion with the patient of the complication and the remedial measures taken helps put the event in perspective and may reduce the likelihood of a legal action. If appropriate, the information should be passed on to the insurance company indemnifying the surgeon and their opinion sought at an early stage.

In conclusion, breakage of instruments is rare during laparoscopic surgery but may occur with thin, sharp instruments introduced percutaneously. Periodic inspection of delicate instruments and their careful maintenance and use should help guard against such mishaps.



 
 ¤ References Top

1.Nathanson LK, Cuschieri A. The falciform lift: a simple method for retraction of the falciform ligament during laparoscopic cholecystectomy. Surg Endosc 1990; 4:186.   Back to cited text no. 1  [PUBMED]  
2.Clarke MT, Arora A, Villar RN. Hip arthroscopy: complications in 1054 cases. Clin Orthop 2003;406:84-8.  Back to cited text no. 2  [PUBMED]  
3.Reed A, Dysert B. Surgical instrument repair. Art or science? Accessed July 1, 2004. http://www.mobileinstrument.com/Articles/Surgical_Instrument_Repair.htm   Back to cited text no. 3    
4.Salameh JR. Suture passer tip breakage during laparoscopic ventral hernia repair. Surg Laparosc Endosc Percutan Tech 2005;15:112-4.  Back to cited text no. 4  [PUBMED]  [FULLTEXT]
5.Lynch CM, Powers AK. Management of a broken needle at the time of laparoscopic Burch. JSLS 2000;4:275-6.  Back to cited text no. 5  [PUBMED]  
6.Kandioler-Eckersberger D, Niederle B, Herbst F, Wenzl E. A magnetic probe to retrieve broken metallic parts of instruments during laparoscopic procedures. Surg Endosc 2002;16:208-9.  Back to cited text no. 6  [PUBMED]  [FULLTEXT]
7.Interpretive guidelines for reporting unusual events. Accessed July 1, 2004. http://www2.state.tn.us/health/Downloads/g4010185.pdf.   Back to cited text no. 7    